Review Article |

Home Blood Pressure Monitoring FREE

Steven A. Yarows, MD; Stevo Julius, MD, ScD; Thomas G. Pickering, MD, DPhil
[+] Author Affiliations

From the Division of Hypertension, University of Michigan Health System, Ann Arbor (Drs Yarows and Julius); and the Department of Medicine, Cardiovascular Center, Cornell Medical Center, New York, NY (Dr Pickering). Dr Yarows has been and is currently a consultant to Omron Healthcare Inc, Vernon Hills, Ill.


Arch Intern Med. 2000;160(9):1251-1257. doi:10.1001/archinte.160.9.1251.
Text Size: A A A
Published online

Hypertension is estimated to affect 43 to 56 million adults or 24% to 31% of the US population and is emerging as a major health problem in some countries in the Third World.13 Hypertension contributes to all the major atherosclerotic cardiovascular disease outcomes.

Blood pressure (BP) can be monitored at the office (clinic) and at home using conventional or ambulatory BP monitors (ABPMs). Ownership of home BP (HBP) monitors is becoming popular and usually occurs without physician encouragement.4 The world BP monitor sales have grown from $484 million in 1992 to $525 million in 1995 and are projected to grow to $597 million by 2002 (Eric Vennemeyer, written communication, December 10, 1997). Higher sales are likely due to the following factors: marketing and a population that is aging and increasingly obese and health conscious. These factors become more important as underdeveloped countries become similar to industrialized nations.5

In the early 1980s, 7.5% to 17.0% of homes in the areas of Minneapolis–St Paul, Minn, and Hamburg, Germany, had sphygmomanometers, whereas 44% to 66% of individuals with hypertension measured their own BP.4,6,7 Of the patients recording the BP at home, 50% to 73% bought the monitor without physician advice and were not trained by qualified personnel.

Despite the common use of these devices by patients, the only recommendation about the correct use of HBP monitors are from ad hoc panels of the American Society of Hypertension8 and the Canadian Coalition for High Blood Pressure Prevention and Control.9 Other groups encourage self-measurement of BP; however, they are not specific on the use of the monitors for assessment of therapy.1014 The British Hypertension Society15 did not mention HBP monitoring in their recommendations for management of essential hypertension. In this article, we review the history of HBP monitoring, types of monitors, how BP is determined, reliability and predictability of HBP monitors, and how HBP monitoring affects outcomes.

The first BP measuring device was developed by the Italian physician Scipione Riva-Rocci, MD, and brought to the United States by Harvey Cushing, MD, in 1901.16 Brown17 in 1930 was the first to report patient self-measurement of systolic BP for assessment of the effect of drugs. Ayman and Goldshine18 were the first to report in 1940 that HBP was less than office BP (OBP). They suggested that HBP monitoring was useful for (1) instructing the patients about their chronic disease, (2) teaching physicians about the natural course of the disease and about factors that affect the disease, (3) learning the prognosis of the disease, and (4) increasing the precision of determining the effectiveness of treatment.

Several types of BP monitors are available for public use. The classic mercury manometer is not popular in the home and is becoming less popular for office use because of environmental concerns about mercury leakage and waste. Aneroid manometers are used with a stethoscope and are usually inexpensive. However, they may be less accurate over time, leading to falsely low readings, and recalibration needs to be performed by the manufacturer.19

Electronic devices may use an auscultatory or oscillometric method of detecting BP. The automated auscultatory method was first developed in the mid-1970s, and the oscillometric method was introduced in 1984.20 The oscillometric method uses the small oscillations in cuff pressure to identify the systolic, mean, and diastolic pressures.21 The mean BP is determined at the peak of the amplitude of the oscillations; the systolic BP, approximately 55% prior to the maximum; and the diastolic BP, approximately 85% after the maximum oscillations, although the exact points are proprietary to each manufacturer (Figure 1). Generally, there is a high correlation between auscultatory and oscillometric devices and simultaneous physician readings.22 The auscultatory method tends to be influenced by unrelated noise, whereas movement influences the oscillometric method. Terminal digit preference, whereby the last numerical digit is rounded, was a more common feature with the stethoscope or aneroid technique (32.3%) compared with the automated electronic monitor (10.5%).23 The oscillometric technique permits faster measurements and is cheaper to manufacture.20 Oscillatory electronic devices have generally replaced auscultatory devices.

Place holder to copy figure label and caption

Oscillometric technique compared with the auscultatory technique for measurement of blood pressure (BP). As indicates auscultatory systolic BP; Am, auscultatory mean BP; and Ad, auscultatory diastolic BP. Reproduced with permission from Biomedical Instrumentation and Technology.21

Graphic Jump Location

Some electronic machines will inflate, deflate, and record the BP automatically, and others need manual inflation and deflation while the monitor records the BP. The self-inflating automated devices are especially useful for patients with arthritis. Some electronic devices have a printer attached and some will store readings that can be downloaded later. Some newer electronic machines use fuzzy logic, which anticipates the systolic BP to allow less overinflation and deflates in a linear as opposed to the usual stepwise fashion. The expense of the devices tends to be related to the degree of automation.

Consumers generally prefer machines with digital readouts and are purchasing an increasing number of electronic HBP monitors (Eric Vennemeyer, written communication, December 10, 1997).24 Sales of electronic monitors have increased from 57% of the world market in 1992 to 58% in 1995 and are expected to increase to 68% in 2002. US consumers purchased 44% of the world market of digital sphygmomanometers in 1992. Finger BP monitors are not accurate for home use.2528

Wrist measurement of BP using the oscillometric technique has been studied in more than 600 adults in 6 studies and in children and adolescents in 1 study (unpublished data from Omron Healthcare Inc, Vernon Hills, Ill).2934 There was generally good agreement compared with the upper arm measurement using auscultatory or oscillometric methods; however, the technique showed high variability in individual cases and the correlation coefficients generally were not as high as those using the upper arm method. The wrist measurements are very dependent on position of the device compared with the heart level. In the standing position with an upper arm cuff, BP is approximately 8 mm Hg higher with the arm positioned by the side rather than at heart level for both systolic and diastolic readings.35,36 The formula for correcting the BP based on arm position is to subtract 2 mm Hg from the reading for every inch (2.54 cm) that the BP cuff is below the horizontal plane of the heart.37 When the wrist manometers are used at the correct measurement level, the measurements are accurate.

Stationary machines are often found in pharmacies for public use. The accuracy may vary considerably between machines. The Vita-Stat automatic, coin-operated BP measurement device (Spacelabs Medical Inc, Redmond, Wash) was found inaccurate for clinical use.3841 The accuracy of other devices that are currently used in community pharmacies has not been determined. These machines may increase the public perception about hypertension and may be considered only for initial self-screening for hypertension.

The common criticism of HBP monitoring is the uncertainty about whether the data are accurate. The reported accuracy of HBP monitors varies. The BP inaccuracy may stem from the operator of the device or from the device itself. There are many reasons within each of these categories, some of which are unique to the device (Table 1). Occasional differences of less than 5 mm Hg are rarely clinically important, especially if many measurements are taken. However, consistent differences of 5 mm Hg may result in the false diagnosis of new or uncontrolled hypertension. Unfortunately, the aneroid sphygmomanometers are similarly unreliable and were found to be inaccurate in 80% of the units tested at a university hospital and clinics.42

The industry is regulated by the Food and Drug Administration in the United States, which uses the Association for the Advancement of Medical Instrumentation (AAMI) standards43 for sphygmomanometers. Other countries use the standards of the British Hypertension Society44,45 for BP measurement. The AAMI standards define passing as 95% or greater of the average measurements agreeing within ±10 mm Hg and 85% agreement within ±5 mm Hg. The device is expected to maintain the safety and performance characteristics for 10,000 full cycles. The pressure indicator accuracy is tested in the range of 20 to 250 mm Hg, and the monitor should not differ by more than ±3 mm Hg or 2% of the readings, whichever is greater. The British Hypertension Society standards assess differences between the test and standard sphygmomanometer by grading (eg, "A" for the most accurate and "D" for the least accurate) based on a summation of the percentage of readings with differences of ±5, ±10, and ±15 mm Hg.

Before accuracy standards were developed, HBP monitoring was usually found to be inaccurate; however, some reports confirming accuracy were also found.7,4648 Once guidelines were published, reports assessed the compliance with the guidelines. Of 23 monitors, 12 (52%) met AAMI standards for diastolic accuracy (39% of devices met both systolic and diastolic standards), whereas 2 of 7 HBP monitors in another study24 failed to meet AAMI criteria. Of the 7, 5 devices (71%) passed an interdevice variability assessment program (conducted by the British Hypertension Society44,45) after 1 month of HBP monitor use twice a day. The following monitors were accurate and complied with the standards of the AAMI43 and/or the British Hypertension Society44,45: Assure models A30 and W20 (Becton Dickinson & Co, Franklin Lakes, NJ); A & D models UA-767 and UA-767PC (A & D Co, Ltd, Tokyo, Japan); Instromedix QD (ALARIS Medical Systems, Inc, San Diego, Calif); Omron models HEM-706, 705CP, 711, 722c, 713c, and 403c (Omron Healthcare Inc); and Terumo model ES-H51 (Terumo Medical Corp, Somerset, NJ) (written communication, Osamu Shirasaki, September 2, 1998; written communication, S. P. Kerestan, July 1998; written communication, Jerry Wang, 1998).23,4954

Home BP monitors have been shown to have reproducible readings, with an SD of less than 3.1 mm Hg for both systolic and diastolic measurements using an artificial oscillometric simulator that eliminates human variation.55 When 2 different machines of the same model were also tested using this same simulator, differences of less than 3.6 mm Hg and SDs less than 0.7 mm Hg for both systolic and diastolic readings were seen. Thus, the results of the electronic machines are reproducible, and variation between machines of the same model is also clinically acceptable.

Although OBP measurement has been the criterion standard, there have been errors with this method in nonresearch settings. The difference between the BP recorded using the correct technique and subsequent office measurements was 6 mm Hg for systolic and 10 mm Hg for diastolic BP.56 Differences between a trained registered nurse and the usual health care providers were 6 mm Hg for systolic and 5 mm Hg for diastolic BP in a family practice center.57 Potentially correctable errors that could be improved by training were estimated to be only 2 mm Hg for systolic and 1 mm Hg for diastolic BP.

The OBP is the criterion standard for determining hypertension-related cardiovascular morbidity and mortality. It is therefore important to establish if HBP measurements are different than OBP measurements. Table 2 shows the summary of the difference between HBP and OBP. On average, OBP is 8.1/5.6 mm Hg higher than HBP.5866

Table Graphic Jump LocationTable 2. The Difference Between Blood Pressure (BP) Measured in the Office and at Home*

Data from ABPM are often reported as 24-hour average and daytime average. The daytime average probably correlates better with the results from HBP monitors since patients usually obtain their BP during the daytime. Table 3 shows that the average difference between the results of daytime ABPM and HBP was −1.7/1.2 mm Hg, which is not clinically important.23,59,60,64,6770

Table Graphic Jump LocationTable 3. The Difference Between Daytime Blood Pressure (BP) Measured by an Ambulatory BP Monitor and at Home*

There are 2 methods to define the limits of normality.71 The first method is the distributional criterion, in which subjects whose BP exceeds 2 SDs of the mean are defined as hypertensive, and the second method relates BP to morbidity and mortality outcomes.

Mejia et al72 reported in 1990 the normal values for HBP based on a population-based study of 608 healthy adults aged 18 to 41 years. Using 2 SDs above the mean of the whole population as a cutoff point, the definition of hypertension for men was a BP of 142/92 mm Hg and for women, 131/85 mm Hg. This population was followed up for a mean of 3 years.73 The authors sought an average HBP that would predict hypertensive BP in future office visits after 3 years with maximum sensitivity and at least 90% specificity. An HBP of 128/83 mm Hg or higher detected sustained hypertension with a sensitivity of 48% and a specificity of 93%.

The PAMELA (Pressione Arteriose Monitorate E Loro Associazioni) study59,60 was an Italian population-based study that followed up 2400 patients aged 25 to 64 years. The accepted upper limit of normal for OBP (140/90 mm Hg) using regression lines corresponded to an upper limit of normal of 120/77 mm Hg for both 24-hour average BP and HBP. The PAMELA project also followed up 400 elderly patients (aged 65 to 74 years).61 Using the regression lines relating HBP to OBP, the upper limit of normal for HBP corresponding to an OBP of 140/90 mm Hg was calculated to be 133/82 mm Hg (95% confidence interval [CI], 131/80-135/83).

A second population-based study67 of 328 subjects determined that the 95th percentile for men aged 20 to 49 years was 150/93 mm Hg and for men aged 50 years and older, it was 157/90 mm Hg. The 95th percentile for women aged 20 to 49 years was 132/82 mm Hg and for women aged 50 years and older, it was 152/87 mm Hg. A third population-based study62 determined that an OBP of 140/90 mm Hg corresponded to an HBP of 127/83 mm Hg. The normal HBP for men was 128/84 mm Hg and for women, 126/83 mm Hg.

Home BP readings were obtained for 1913 population-based subjects who were followed up for a mean of 5 years in a rural Japanese community.74 An "excess of relative risk" of 10% was arbitrarily accepted as a serious and substantial risk. Based on this assumption, the home reference value for defining hypertension was calculated to be 137/84 mm Hg or higher, which was equivalent to an excess risk associated with an OBP of 140/90 mm Hg or higher. The relative hazard (RH) of home systolic BP of 137 mm Hg or higher was 1.77 and of diastolic BP of 84 mm Hg or higher, 1.70. These investigators found that a U-shaped curve and a diastolic BP less than 65 mm Hg were associated with an increase of the RH to 1.54.

Despite the differences in methods, the data across studies are reasonably consistent and suggest that the upper limit of normal HBP should be 130/80 to 135/85 mm Hg. This result is comparable to that reported in a meta-analysis75 of 17 previous studies that suggested an upper limit of normal of 135/85 mm Hg.

End-diastolic relative wall thickness was measured using M-mode echocardiograms in a comparative study of home, office, and 24-hour ambulatory BP monitoring.64 The correlation coefficients were higher with home (r=0.45/0.40; P<.01) than office (r=0.22/0.07) or 24-hour ambulatory BP (r=0.26/0.24) readings.

The effect of BP control on the evolution of electrocardiographic evidence of left ventricular hypertrophy was studied in 50 patients during an average 9-year follow-up.76 The changes in electrocardiographic evidence of hypertrophy were related to the degree of BP control and correlated better with HBP average measurements rather than office measurements.

Home BP readings were obtained in a population-based study74 for 1913 subjects whose mean age was 61 years who underwent a mean follow-up of 5 years in a rural Japanese community. The predictive power of HBP levels for subsequent mortality was slightly stronger than that of OBP. The RH for overall mortality for systolic HBP higher than 138 mm Hg was 2.06 (95% CI, 1.18-3.58; P<.05), whereas for systolic OBP of 140 mm Hg or higher, the RH was 1.50 (95% CI, 1.05-2.15). The RH of diastolic HBP higher than 83 mm Hg was 1.90 (95% CI, 1.12-3.24; P<.05), with a comparative RH for OBP higher than 90 mm Hg of 1.06 (95% CI, 0.62-1.79).

Approximately 60% of the patients thought that HBP monitoring enabled them to cope better with the disease.77 Physicians thought that patients who used HBP monitoring and had elevated BP values "reacted on the whole reasonably," and the use of HBP monitoring allowed the patients to have a positive attitude toward their disease.7 Of patients in one study,78 96% thought that measuring their BP at home was "worthwhile" because of the reassurance that their BP was controlled.

The reports in the literature are mixed regarding improvement in BP control using HBP monitoring. Home BP monitoring was found not to lower BP after 6 months to 1 year of use in studies of patients with mild hypertension and patients with difficult to control hypertension.79,80 Home BP monitoring was shown to lower BP in other studies (including one randomized, controlled trial)78,81,82 with patients monitored for 6 months to 1 year. When patient self-directed adjustment of medication was combined with HBP monitoring, there was a significant decrease in mean BP measured by ABPM (−0.95 mm Hg) after 8 weeks compared with the office-based, standard management group (+1.90 mm Hg; P=.04).83 There is enough evidence of HBP monitoring improving BP control to justify a large randomized long-term study to clarify this issue.

The reports in the literature are also mixed regarding the success of using HBP monitoring to improve compliance with therapy. Variations in study designs and definitions of compliance make it difficult to reach a definite conclusion; however, the general trend suggests a benefit.78,8387

There are costs in measuring BP using office visits with or without the physician present. The yearly ambulatory care cost of hypertension treatment per patient at Veterans Affairs hypertension clinics was estimated as $647 in 1989 dollars.88 Forty-nine percent of this cost was for clinic visits. We estimated that the total direct cost in Ann Arbor, Mich, for simply measuring BP in an office, without a physician present, is approximately $5.26. Patients also incur direct and indirect costs.

In a study by Soghikan et al,78 there were 1.2 (95% CI, −1.7 to −0.8) fewer physician visits over a year in a home-monitored group compared with the group that received usual care. The mean cost of hypertension care per patient in the home-monitored group was $117, which is 6% less than the group that received usual care. Although this difference was not significant (P=.44), out-of-pocket costs for transportation and lost wages were not calculated.

Home BP monitoring has become accurate since standards were developed. The results of monitoring are reproducible, with differences between models less than the differences due to human variation in the auscultation of BP. Home measurements result in lower BP readings than office measurements, with normal values similar to daytime ABPM. An OBP of 140/90 mm Hg is approximately equivalent to an HBP of 130/84 mm Hg. Although several studies indicate that HBP better predicts left ventricular hypertrophy and mortality than OBP, further research is needed. Additional studies are also warranted to determine if HBP monitoring improves BP control and compliance and is cost-effective, although there have been studies showing benefits in each of these areas. It may be time to transfer HBP monitoring from a consumer market to a physician-endorsed activity, with appropriate third-party reimbursement for and physician review of the monitors.

Accepted for publication September 9, 1999.

Reprints: Steven A. Yarows, MD, Division of Hypertension, University of Michigan Health System, 128 Van Buren, Chelsea, MI 48118 (e-mail: syarows@umich.edu).

Burt  VLWhelton  PRoccella  EJ  et al.  Prevalence of hypertension in the US adult population: results from the Third National Health and Nutrition Examination Survey, 1988-1991. Hypertension. 1995;25305- 313
Gupta  R Meta-analysis of prevalence of hypertension in India. Indian Heart J. 1997;49450
Ramirez  MOPino  CTFuriasse  LVLee  AJFowkes  FG Paraguayan National Blood Pressure Study: prevalence of hypertension in the general population. J Hum Hypertens. 1995;9891- 897
Jeck  TEdmonds  DMengden  TSchubert  MVetter  W Performing self-measurement of blood pressure: a patient survey [in German]. Schweiz Rundsch Med Prax. 1991;80456- 461
Not Available, Frost & Sullivan information page. Profound Web site Available at: http://www.profound.com/info/sources/researchline/f&s.shtml. Accessed February 23, 2000
Hahn  LPFolsom  ARSprafka  JMPrineas  RJ Prevalence and accuracy of home sphygmomanometers in an urban population. Am J Public Health. 1987;771459- 1461
Krecke  HJFleischmann  CBokmann  M Distribution and acceptance of self-measurement of blood pressure in the Hamburg area [in German]. Schweiz Rundsch Med Prax. 1989;781336- 1342
Pickering  T Recommendations for the use of home (self) and ambulatory blood pressure monitoring. Am J Hypertens. 1995;91- 11
Campbell  NRAbbott  DBass  M  et al.  Self-measurement of blood pressure: recommendations of the Canadian Coalition for High Blood Pressure Prevention and Control. Can J Cardiol. 1995;11 ((suppl H)) 5H- 17H
Not Available, The Sixth Report of the Joint National Committee on Detection, Evaluation, and Treatment of High Blood Pressure. Arch Intern Med. 1997;1572413- 2446
American College of Physicians, Automated ambulatory blood pressure and self-measured blood pressure monitoring devices: their role in the diagnosis and management of hypertension. Ann Intern Med. 1993;118889- 892
Not Available, 1986 Guidelines for the treatment of mild hypertension: memorandum from a WHO/ISH meeting. J Hypertens. 1986;4383- 386
Not Available, The 1988 Report of the Joint National Committee on Detection, Evaluation, and Treatment of High Blood Pressure. Arch Intern Med. 1988;1481023- 1038
Not Available, The Fifth Report of the Joint National Committee on Detection, Evaluation, and Treatment of High Blood Pressure. Arch Intern Med. 1993;153154- 183
Sever  PBeevers  GBulpitt  C  et al.  Management guidelines in essential hypertension: report of the second working party of the British Hypertension Society. BMJ. 1993;306983- 987
Crenner  CW Introduction of the blood pressure cuff into U.S. medical practice: technology and skilled practice. Ann Intern Med. 1998;128488- 493
Brown  GE Daily and monthly rhythm in the blood pressure of a man with hypertension: a three-year study. Ann Intern Med. 1930;31177- 1189
Ayman  DGoldshine  AD Blood pressure determinations by patients with essential hypertension, I: the difference between clinic and home readings before treatment. Am J Med Sci. 1940;200465- 474
Petrie  JCO'Brien  ETLittler  WAde Swiet  M Recommendations on blood pressure measurements. BMJ (Clin Res Ed). 1986;293611- 615
Mieke  S Substitution of simulators for human subjects. Blood Press Monit. 1997;2251- 256
Yong  PGeddes  LA A surrogate arm for evaluating the accuracy of instruments for indirect measurement of blood pressure. Biomed Instrum Technol. 1990;24130- 135
Zachariah  PKSheps  SGSmith  RL Role of self-monitoring and ambulatory monitoring in diagnosis and evaluation of hypertension. Practical Cardiol. 1988;141- 7
Stergiou  GSVoutsa  AVAchimastos  ADMountokalakis  TD Home self-monitoring of blood pressure: is fully automated oscillometric technique as good as conventional stethoscopic technique? Am J Hypertens. 1997;10 ((4, pt 1)) 428- 433
Evans  CEHaynes  RBGoldsmith  CHHewson  SA Home blood pressure-measuring devices: a comparative study of accuracy. J Hypertens. 1989;7133- 142
Nesselroad  JMFlacco  VAPhillips  KMKruse  J Accuracy of automated finger blood pressure devices. Fam Med. 1996;28182- 192
Veerman  DPLenders  JWThien  Tvan Montfrans  GA LAM 100/Marshall F-88: accuracy and precision of a new device for discontinuous finger blood pressure measurement. J Hum Hypertens. 1993;7113- 115
Teshima  MKuwajima  IInukai  MSuzuki  YMatsusita  SKuramoto  K Clinical evaluation of finger blood pressure measurement devices for home-use [in Japanese]. Nippon Ronen Igakkai Zasshi. 1993;3054- 58
Iyriboz  Y Oscillometric finger blood pressure versus brachial auscultative blood pressure recording. J Fam Pract. 1990;31376- 380
Weber  FErbel  R Measuring blood pressure at the wrist: critical analysis of a validation study [in German]. Med Klin. 1995;90562- 566
Bald  MWesthues  RBonzel  KE Blood pressure monitoring at the wrist: is it reliable in children and adolescents? Z Kardiol. 1996;85 ((suppl 3)) 106- 108
Thummler  MWonka  FSchoppe  A Preliminary clinical comparative study of a new blood pressure instrument with wrist cuff [in German]. Z Kardiol. 1994;83641- 645
Eckert  SGleichmann  SGleichmann  U Blood pressure self-measurement in upper arm and in wrist for treatment control of arterial hypertension compared to ABPM. Z Kardiol. 1996;85 ((suppl 3)) 109- 111
Saul  FKlaus  DAristidou  YWiemeyer  ALosse  B Non-invasive oscillometric wrist and upper arm blood pressure measurements compared with invasive values. Z Kardiol. 1996;85127- 129
Latman  NSLatman  A Evaluation of instruments for noninvasive blood pressure monitoring of the wrist. Biomed Instrum Technol. 1997;3163- 68
Mariotti  GAlii  CAvanzini  F  et al.  Arm position as a source of error in blood pressure measurement. Clin Cardiol. 1987;10591- 593
Waal-Manning  HJPaulin  JM Effects of arm position and support on blood pressure readings. J Clin Hypertens. 1987;3624- 630
Ramsey  M Blood pressure monitoring: automated oscillometric devices. J Clin Monit. 1991;756- 67
Polk  BFRosner  BFeudo  RVandenburgh  M An evaluation of the Vita-Stat automatic blood pressure measuring device. Hypertension. 1980;2221- 227
Whelton  PKThompson  SGBarnes  GRMiall  WE Evaluation of the Vita-Stat automatic blood pressure recorder: a comparison with the Random-Zero sphygmomanometer. Am J Epidemiol. 1983;11746- 54
Salaita  KWhelton  PKSeidler  AJ A community-based evaluation of the Vita-Stat automatic blood pressure recorder. Am J Hypertens. 1990;3 ((5, pt 1)) 366- 372
Whitcomb  BLProchazka  ALoVerde  MByyny  RL Failure of the community-based Vita-Stat automated blood pressure device to accurately measure blood pressure. Arch Fam Med. 1995;4419- 424
Bailey  RHKnaus  VLBauer  JH Aneroid sphygmomanometers: an assessment of accuracy at a university hospital and clinics. Arch Intern Med. 1991;1511409- 1412
Association for the Advancement of Medical Instrumentation, Electronic or Automated Sphygmomanometers.  Arlington, Va American National Standards Institute Inc1992;1- 40
O'Brien  EPetrie  JLittler  W  et al.  The British Hypertension Society protocol for the evaluation of automated and semi-automated blood pressure measuring devices with special reference to ambulatory systems. J Hypertens. 1990;8607- 619
O'Brien  EPetrie  JLittler  W  et al.  Short report: an outline of the revised British Hypertension Society protocol for the evaluation of blood pressure measuring devices. J Hypertens. 1993;11677- 679
Weber  FHirche  HSimonides  RAnlauf  M Comparative clinical study on the accuracy of conventional and electronic blood pressure self-measuring devices [in German]. Z Kardiol. 1981;70700- 705
Harrison  DWKelly  PL Home health-care: accuracy, calibration, exhaust, and failure rate comparisons of digital blood pressure monitors. Med Instrum. 1987;21323- 328
Imai  YAbe  KSasaki  S  et al.  Clinical evaluation of semiautomatic and automatic devices for home blood pressure measurement: comparison between cuff-oscillometric and microphone methods. J Hypertens. 1989;7983- 990
James  GDYee  LSCates  EMSchlussel  YRPecker  MSPickering  TG A validation study of the Instromedix Baro-Graf QD home blood pressure monitor. Am J Hypertens. 1990;3717- 720
Foster  CMcKinlay  SCruickshank  JMCoats  AJS Accuracy of the Omron HEM 706 portable monitor for home measurement of blood pressure. J Hum Hypertens. 1994;8661- 664
Cordoba  RFuertes  MIAlvarez  AMolina  ISolans  RMelero  I The evaluation of a self-measurement arterial pressure monitor: the OMRON-HM 722C [in Spanish]. Aten Primaria. 1997;20247- 250
Mufunda  JSparks  BChifamba  J  et al.  Comparison of the Omron HEM-713C automated blood pressure monitor with a standard ausculatory method using a mercury manometer. Cent Afr J Med. 1996;42230- 232
Walma  EPvan Dooren  Cvan der Does  EPrins  AMulder  PHoes  AW Accuracy of an oscillometric automatic blood pressure device: the Omron HEM403C. J Hum Hypertens. 1995;9169- 174
Kwek  KChan  YTan  KHYeo  G Validation of an oscillometric electronic sphygmomanometer in an obstetric population. Am J Hypertens. 1998;11978- 982
Yarows  SAAmerena  JV Accuracy of 10 models of home blood pressure monitors using a oscillometric simulator. Blood Press Monit. 1999;445- 52
Rocha  JCRocha  ATMagossi  AMGLeao  RWPalu  MJFMoreira  DC Evaluation of the technique for taking blood pressure by health care workers in an university hospital. Am J Hypertens. 1988;1166AAbstract
Kay  LE Accuracy of blood pressure measurement in the family practice center. J Am Board Fam Pract. 1998;11252- 258
Welin  LSvärdsudd  KTibblin  G Home blood pressure measurements: feasibility and results compared to office measurements. Acta Med Scand. 1982;211275- 279
Sega  GCesana  GValagussa  FMancia  GZanchetti  A Ambulatory and home blood pressure normality: the PAMELA study. J Cardiovasc Pharmacol. 1994;23 ((suppl 5)) S12- S15
Mancia  GSega  RBravi  C  et al.  Ambulatory blood pressure normality: results from the PAMELA study. J Hypertens. 1995;13 ((12, pt 1)) 1377- 1390
Sega  RBravi  CCesana  GMilesi  CGrassi  GZanchetti  AMancia  G Ambulatory and home blood pressure normality in the elderly: data from the PAMELA population. Hypertension. 1997;30 ((1, pt 1)) 1- 6
de Gaudemaris  RChau  NPMallion  JM Home blood pressure: variability, comparison with office readings and proposal for reference values. J Hypertens. 1994;12831- 838
Appel  LJStason  WB Ambulatory blood pressure monitoring and blood pressure self-measurement in the diagnosis and management of hypertension. Ann Intern Med. 1993;118867- 882
Kleinert  HDHarshfield  GAPicketing  TG  et al.  What is the value of home blood pressure measurement in patients with mild hypertension? Hypertension. 1984;6574- 578
Bobrie  GDay  MTugayé  AChatellier  GMonad  J Self blood pressure measurement at home. Clin Exp Hypertens. 1993;151109- 1119
Battig  BSteiner  AJeck  TVetter  W Blood pressure self-measurement in normotensive and hypertensive patients. J Hypertens Suppl. 1989;7S59- S63
Staessen  JBulpitt  CJFagard  R  et al.  Reference values for the ambulatory blood pressure and the blood pressure measured at home: a population study. J Hum Hypertens. 1991;5355- 361
Staessen  JAFagard  RLijnen  P  et al.  Ambulatory blood pressure and blood pressure measured at home: progress report on a population study. J Cardiovasc Pharmacol. 1994;23 ((suppl 5)) S5- S11
Bialy  GBRuddy  MCMalka  ESSilvay  LAKamalakannan  N Comparison of office, home and 24-hour ambulatory blood pressures in borderline and mild hypertension. Angiology. 1988;39752- 760
Stergiou  GSMalakos  JSVoutsa  AVAchimastos  ADMountokalakis  TD Home monitoring of blood pressure: limited value in general practice. J Hum Hypertens. 1996;10219- 223
Mejia  AJulius  S Practical utility of blood pressure readings obtained by self-determination. J Hypertens Suppl. 1989;7 ((suppl 3)) S53- S57
Mejia  ADJulius  SJones  KASchork  NJKneisley  J The Tecumseh Blood Pressure Study: normative data on blood pressure self-determination. Arch Intern Med. 1990;1501209- 1213
Nesbitt  SDAmerena  JVGrant  E  et al.  Home blood pressure as a predictor of future blood pressure stability in borderline hypertension: the Tecumseh Study. Am J Hypertens. 1997;101270- 1280
Tsuji  IImai  YNagai  K  et al.  Proposal of reference values for home blood pressure measurement: prognostic criteria based on a prospective observation of the general population in Ohasama, Japan. Am J Hypertens. 1997;10 ((4, pt 1)) 409- 418
Thijs  LStaessen  JACelis  H  et al.  Reference values for self-recorded blood pressure: a meta-analysis of summary data. Arch Intern Med. 1998;158481- 488
Ibrahim  MMTarazi  RCDustan  HPGifford  RW  Jr Electrocardiogram in evaluation of resistance to antihypertensive therapy. Arch Intern Med. 1977;1371125- 1129
Krecke  HJLutkes  PMaiwald  MSchultze-Rupp  A Self-measurement of blood pressure in hypertensive subjects in Germany: results of a questionnaire in spring/early summer 1993 [in German]. Schweiz Rundsch Med Prax. 1994;83895- 900
Soghikan  KCasper  SMFireman  BH  et al.  Home blood pressure monitoring: effect on use of medical services and medical care costs. Med Care. 1992;30855- 865
Johnson  ALTaylor  DWSackett  DLDunnett  CWShimizu  AG Self-recording of blood pressure in the management of hypertension. Can Med Assoc J. 1978;1191034- 1039
Midanik  LTResnick  BHurley  LBSmith  EJMcCarthy  M Home blood pressure monitoring for mild hypertensives. Public Health Rep. 1991;10685- 89
Stahl  SMKelley  CRNeill  PJGrim  CEMamlin  J Effects of home blood pressure measurement on long-term control. Am J Public Health. 1984;74704- 709
Carnahan  JENugent  CA The effects of self-monitoring by patients on the control of hypertension. Am J Med Sci. 1975;26969- 73
Zarnke  KBFeagan  BGMahon  JLFeldman  RD A randomized study comparing a patient-directed hypertension management strategy with usual office-based care. Am J Hypertens. 1997;1058- 67
Midanik  LTResnick  BHurley  LBSmith  EJMcCarthy  M Home blood pressure monitoring in hypertension. Public Health Rep. 1991;10685- 89
Edmonds  DFoerster  EGroth  HGreminger  PSiegenthaler  WVetter  W Does self-measurement of blood pressure improve patient compliance in hypertension? J Hypertens Suppl. 1985;3S31- S34
Haynes  RBSackett  DLGibson  ES  et al.  Improvement of medication compliance in uncontrolled hypertension. Lancet. 1976;11265- 1268
Nessman  DGCarnahan  JENugent  CA Increasing compliance: patient-operated hypertension groups. Arch Intern Med. 1980;1401427- 1430
Stason  WBShepard  DSPerry  M  et al.  Effectiveness and costs of Veterans Affairs hypertension clinics. Med Care. 1994;321197- 1215

Figures

Place holder to copy figure label and caption

Oscillometric technique compared with the auscultatory technique for measurement of blood pressure (BP). As indicates auscultatory systolic BP; Am, auscultatory mean BP; and Ad, auscultatory diastolic BP. Reproduced with permission from Biomedical Instrumentation and Technology.21

Graphic Jump Location

Tables

Table Graphic Jump LocationTable 2. The Difference Between Blood Pressure (BP) Measured in the Office and at Home*
Table Graphic Jump LocationTable 3. The Difference Between Daytime Blood Pressure (BP) Measured by an Ambulatory BP Monitor and at Home*

References

Burt  VLWhelton  PRoccella  EJ  et al.  Prevalence of hypertension in the US adult population: results from the Third National Health and Nutrition Examination Survey, 1988-1991. Hypertension. 1995;25305- 313
Gupta  R Meta-analysis of prevalence of hypertension in India. Indian Heart J. 1997;49450
Ramirez  MOPino  CTFuriasse  LVLee  AJFowkes  FG Paraguayan National Blood Pressure Study: prevalence of hypertension in the general population. J Hum Hypertens. 1995;9891- 897
Jeck  TEdmonds  DMengden  TSchubert  MVetter  W Performing self-measurement of blood pressure: a patient survey [in German]. Schweiz Rundsch Med Prax. 1991;80456- 461
Not Available, Frost & Sullivan information page. Profound Web site Available at: http://www.profound.com/info/sources/researchline/f&s.shtml. Accessed February 23, 2000
Hahn  LPFolsom  ARSprafka  JMPrineas  RJ Prevalence and accuracy of home sphygmomanometers in an urban population. Am J Public Health. 1987;771459- 1461
Krecke  HJFleischmann  CBokmann  M Distribution and acceptance of self-measurement of blood pressure in the Hamburg area [in German]. Schweiz Rundsch Med Prax. 1989;781336- 1342
Pickering  T Recommendations for the use of home (self) and ambulatory blood pressure monitoring. Am J Hypertens. 1995;91- 11
Campbell  NRAbbott  DBass  M  et al.  Self-measurement of blood pressure: recommendations of the Canadian Coalition for High Blood Pressure Prevention and Control. Can J Cardiol. 1995;11 ((suppl H)) 5H- 17H
Not Available, The Sixth Report of the Joint National Committee on Detection, Evaluation, and Treatment of High Blood Pressure. Arch Intern Med. 1997;1572413- 2446
American College of Physicians, Automated ambulatory blood pressure and self-measured blood pressure monitoring devices: their role in the diagnosis and management of hypertension. Ann Intern Med. 1993;118889- 892
Not Available, 1986 Guidelines for the treatment of mild hypertension: memorandum from a WHO/ISH meeting. J Hypertens. 1986;4383- 386
Not Available, The 1988 Report of the Joint National Committee on Detection, Evaluation, and Treatment of High Blood Pressure. Arch Intern Med. 1988;1481023- 1038
Not Available, The Fifth Report of the Joint National Committee on Detection, Evaluation, and Treatment of High Blood Pressure. Arch Intern Med. 1993;153154- 183
Sever  PBeevers  GBulpitt  C  et al.  Management guidelines in essential hypertension: report of the second working party of the British Hypertension Society. BMJ. 1993;306983- 987
Crenner  CW Introduction of the blood pressure cuff into U.S. medical practice: technology and skilled practice. Ann Intern Med. 1998;128488- 493
Brown  GE Daily and monthly rhythm in the blood pressure of a man with hypertension: a three-year study. Ann Intern Med. 1930;31177- 1189
Ayman  DGoldshine  AD Blood pressure determinations by patients with essential hypertension, I: the difference between clinic and home readings before treatment. Am J Med Sci. 1940;200465- 474
Petrie  JCO'Brien  ETLittler  WAde Swiet  M Recommendations on blood pressure measurements. BMJ (Clin Res Ed). 1986;293611- 615
Mieke  S Substitution of simulators for human subjects. Blood Press Monit. 1997;2251- 256
Yong  PGeddes  LA A surrogate arm for evaluating the accuracy of instruments for indirect measurement of blood pressure. Biomed Instrum Technol. 1990;24130- 135
Zachariah  PKSheps  SGSmith  RL Role of self-monitoring and ambulatory monitoring in diagnosis and evaluation of hypertension. Practical Cardiol. 1988;141- 7
Stergiou  GSVoutsa  AVAchimastos  ADMountokalakis  TD Home self-monitoring of blood pressure: is fully automated oscillometric technique as good as conventional stethoscopic technique? Am J Hypertens. 1997;10 ((4, pt 1)) 428- 433
Evans  CEHaynes  RBGoldsmith  CHHewson  SA Home blood pressure-measuring devices: a comparative study of accuracy. J Hypertens. 1989;7133- 142
Nesselroad  JMFlacco  VAPhillips  KMKruse  J Accuracy of automated finger blood pressure devices. Fam Med. 1996;28182- 192
Veerman  DPLenders  JWThien  Tvan Montfrans  GA LAM 100/Marshall F-88: accuracy and precision of a new device for discontinuous finger blood pressure measurement. J Hum Hypertens. 1993;7113- 115
Teshima  MKuwajima  IInukai  MSuzuki  YMatsusita  SKuramoto  K Clinical evaluation of finger blood pressure measurement devices for home-use [in Japanese]. Nippon Ronen Igakkai Zasshi. 1993;3054- 58
Iyriboz  Y Oscillometric finger blood pressure versus brachial auscultative blood pressure recording. J Fam Pract. 1990;31376- 380
Weber  FErbel  R Measuring blood pressure at the wrist: critical analysis of a validation study [in German]. Med Klin. 1995;90562- 566
Bald  MWesthues  RBonzel  KE Blood pressure monitoring at the wrist: is it reliable in children and adolescents? Z Kardiol. 1996;85 ((suppl 3)) 106- 108
Thummler  MWonka  FSchoppe  A Preliminary clinical comparative study of a new blood pressure instrument with wrist cuff [in German]. Z Kardiol. 1994;83641- 645
Eckert  SGleichmann  SGleichmann  U Blood pressure self-measurement in upper arm and in wrist for treatment control of arterial hypertension compared to ABPM. Z Kardiol. 1996;85 ((suppl 3)) 109- 111
Saul  FKlaus  DAristidou  YWiemeyer  ALosse  B Non-invasive oscillometric wrist and upper arm blood pressure measurements compared with invasive values. Z Kardiol. 1996;85127- 129
Latman  NSLatman  A Evaluation of instruments for noninvasive blood pressure monitoring of the wrist. Biomed Instrum Technol. 1997;3163- 68
Mariotti  GAlii  CAvanzini  F  et al.  Arm position as a source of error in blood pressure measurement. Clin Cardiol. 1987;10591- 593
Waal-Manning  HJPaulin  JM Effects of arm position and support on blood pressure readings. J Clin Hypertens. 1987;3624- 630
Ramsey  M Blood pressure monitoring: automated oscillometric devices. J Clin Monit. 1991;756- 67
Polk  BFRosner  BFeudo  RVandenburgh  M An evaluation of the Vita-Stat automatic blood pressure measuring device. Hypertension. 1980;2221- 227
Whelton  PKThompson  SGBarnes  GRMiall  WE Evaluation of the Vita-Stat automatic blood pressure recorder: a comparison with the Random-Zero sphygmomanometer. Am J Epidemiol. 1983;11746- 54
Salaita  KWhelton  PKSeidler  AJ A community-based evaluation of the Vita-Stat automatic blood pressure recorder. Am J Hypertens. 1990;3 ((5, pt 1)) 366- 372
Whitcomb  BLProchazka  ALoVerde  MByyny  RL Failure of the community-based Vita-Stat automated blood pressure device to accurately measure blood pressure. Arch Fam Med. 1995;4419- 424
Bailey  RHKnaus  VLBauer  JH Aneroid sphygmomanometers: an assessment of accuracy at a university hospital and clinics. Arch Intern Med. 1991;1511409- 1412
Association for the Advancement of Medical Instrumentation, Electronic or Automated Sphygmomanometers.  Arlington, Va American National Standards Institute Inc1992;1- 40
O'Brien  EPetrie  JLittler  W  et al.  The British Hypertension Society protocol for the evaluation of automated and semi-automated blood pressure measuring devices with special reference to ambulatory systems. J Hypertens. 1990;8607- 619
O'Brien  EPetrie  JLittler  W  et al.  Short report: an outline of the revised British Hypertension Society protocol for the evaluation of blood pressure measuring devices. J Hypertens. 1993;11677- 679
Weber  FHirche  HSimonides  RAnlauf  M Comparative clinical study on the accuracy of conventional and electronic blood pressure self-measuring devices [in German]. Z Kardiol. 1981;70700- 705
Harrison  DWKelly  PL Home health-care: accuracy, calibration, exhaust, and failure rate comparisons of digital blood pressure monitors. Med Instrum. 1987;21323- 328
Imai  YAbe  KSasaki  S  et al.  Clinical evaluation of semiautomatic and automatic devices for home blood pressure measurement: comparison between cuff-oscillometric and microphone methods. J Hypertens. 1989;7983- 990
James  GDYee  LSCates  EMSchlussel  YRPecker  MSPickering  TG A validation study of the Instromedix Baro-Graf QD home blood pressure monitor. Am J Hypertens. 1990;3717- 720
Foster  CMcKinlay  SCruickshank  JMCoats  AJS Accuracy of the Omron HEM 706 portable monitor for home measurement of blood pressure. J Hum Hypertens. 1994;8661- 664
Cordoba  RFuertes  MIAlvarez  AMolina  ISolans  RMelero  I The evaluation of a self-measurement arterial pressure monitor: the OMRON-HM 722C [in Spanish]. Aten Primaria. 1997;20247- 250
Mufunda  JSparks  BChifamba  J  et al.  Comparison of the Omron HEM-713C automated blood pressure monitor with a standard ausculatory method using a mercury manometer. Cent Afr J Med. 1996;42230- 232
Walma  EPvan Dooren  Cvan der Does  EPrins  AMulder  PHoes  AW Accuracy of an oscillometric automatic blood pressure device: the Omron HEM403C. J Hum Hypertens. 1995;9169- 174
Kwek  KChan  YTan  KHYeo  G Validation of an oscillometric electronic sphygmomanometer in an obstetric population. Am J Hypertens. 1998;11978- 982
Yarows  SAAmerena  JV Accuracy of 10 models of home blood pressure monitors using a oscillometric simulator. Blood Press Monit. 1999;445- 52
Rocha  JCRocha  ATMagossi  AMGLeao  RWPalu  MJFMoreira  DC Evaluation of the technique for taking blood pressure by health care workers in an university hospital. Am J Hypertens. 1988;1166AAbstract
Kay  LE Accuracy of blood pressure measurement in the family practice center. J Am Board Fam Pract. 1998;11252- 258
Welin  LSvärdsudd  KTibblin  G Home blood pressure measurements: feasibility and results compared to office measurements. Acta Med Scand. 1982;211275- 279
Sega  GCesana  GValagussa  FMancia  GZanchetti  A Ambulatory and home blood pressure normality: the PAMELA study. J Cardiovasc Pharmacol. 1994;23 ((suppl 5)) S12- S15
Mancia  GSega  RBravi  C  et al.  Ambulatory blood pressure normality: results from the PAMELA study. J Hypertens. 1995;13 ((12, pt 1)) 1377- 1390
Sega  RBravi  CCesana  GMilesi  CGrassi  GZanchetti  AMancia  G Ambulatory and home blood pressure normality in the elderly: data from the PAMELA population. Hypertension. 1997;30 ((1, pt 1)) 1- 6
de Gaudemaris  RChau  NPMallion  JM Home blood pressure: variability, comparison with office readings and proposal for reference values. J Hypertens. 1994;12831- 838
Appel  LJStason  WB Ambulatory blood pressure monitoring and blood pressure self-measurement in the diagnosis and management of hypertension. Ann Intern Med. 1993;118867- 882
Kleinert  HDHarshfield  GAPicketing  TG  et al.  What is the value of home blood pressure measurement in patients with mild hypertension? Hypertension. 1984;6574- 578
Bobrie  GDay  MTugayé  AChatellier  GMonad  J Self blood pressure measurement at home. Clin Exp Hypertens. 1993;151109- 1119
Battig  BSteiner  AJeck  TVetter  W Blood pressure self-measurement in normotensive and hypertensive patients. J Hypertens Suppl. 1989;7S59- S63
Staessen  JBulpitt  CJFagard  R  et al.  Reference values for the ambulatory blood pressure and the blood pressure measured at home: a population study. J Hum Hypertens. 1991;5355- 361
Staessen  JAFagard  RLijnen  P  et al.  Ambulatory blood pressure and blood pressure measured at home: progress report on a population study. J Cardiovasc Pharmacol. 1994;23 ((suppl 5)) S5- S11
Bialy  GBRuddy  MCMalka  ESSilvay  LAKamalakannan  N Comparison of office, home and 24-hour ambulatory blood pressures in borderline and mild hypertension. Angiology. 1988;39752- 760
Stergiou  GSMalakos  JSVoutsa  AVAchimastos  ADMountokalakis  TD Home monitoring of blood pressure: limited value in general practice. J Hum Hypertens. 1996;10219- 223
Mejia  AJulius  S Practical utility of blood pressure readings obtained by self-determination. J Hypertens Suppl. 1989;7 ((suppl 3)) S53- S57
Mejia  ADJulius  SJones  KASchork  NJKneisley  J The Tecumseh Blood Pressure Study: normative data on blood pressure self-determination. Arch Intern Med. 1990;1501209- 1213
Nesbitt  SDAmerena  JVGrant  E  et al.  Home blood pressure as a predictor of future blood pressure stability in borderline hypertension: the Tecumseh Study. Am J Hypertens. 1997;101270- 1280
Tsuji  IImai  YNagai  K  et al.  Proposal of reference values for home blood pressure measurement: prognostic criteria based on a prospective observation of the general population in Ohasama, Japan. Am J Hypertens. 1997;10 ((4, pt 1)) 409- 418
Thijs  LStaessen  JACelis  H  et al.  Reference values for self-recorded blood pressure: a meta-analysis of summary data. Arch Intern Med. 1998;158481- 488
Ibrahim  MMTarazi  RCDustan  HPGifford  RW  Jr Electrocardiogram in evaluation of resistance to antihypertensive therapy. Arch Intern Med. 1977;1371125- 1129
Krecke  HJLutkes  PMaiwald  MSchultze-Rupp  A Self-measurement of blood pressure in hypertensive subjects in Germany: results of a questionnaire in spring/early summer 1993 [in German]. Schweiz Rundsch Med Prax. 1994;83895- 900
Soghikan  KCasper  SMFireman  BH  et al.  Home blood pressure monitoring: effect on use of medical services and medical care costs. Med Care. 1992;30855- 865
Johnson  ALTaylor  DWSackett  DLDunnett  CWShimizu  AG Self-recording of blood pressure in the management of hypertension. Can Med Assoc J. 1978;1191034- 1039
Midanik  LTResnick  BHurley  LBSmith  EJMcCarthy  M Home blood pressure monitoring for mild hypertensives. Public Health Rep. 1991;10685- 89
Stahl  SMKelley  CRNeill  PJGrim  CEMamlin  J Effects of home blood pressure measurement on long-term control. Am J Public Health. 1984;74704- 709
Carnahan  JENugent  CA The effects of self-monitoring by patients on the control of hypertension. Am J Med Sci. 1975;26969- 73
Zarnke  KBFeagan  BGMahon  JLFeldman  RD A randomized study comparing a patient-directed hypertension management strategy with usual office-based care. Am J Hypertens. 1997;1058- 67
Midanik  LTResnick  BHurley  LBSmith  EJMcCarthy  M Home blood pressure monitoring in hypertension. Public Health Rep. 1991;10685- 89
Edmonds  DFoerster  EGroth  HGreminger  PSiegenthaler  WVetter  W Does self-measurement of blood pressure improve patient compliance in hypertension? J Hypertens Suppl. 1985;3S31- S34
Haynes  RBSackett  DLGibson  ES  et al.  Improvement of medication compliance in uncontrolled hypertension. Lancet. 1976;11265- 1268
Nessman  DGCarnahan  JENugent  CA Increasing compliance: patient-operated hypertension groups. Arch Intern Med. 1980;1401427- 1430
Stason  WBShepard  DSPerry  M  et al.  Effectiveness and costs of Veterans Affairs hypertension clinics. Med Care. 1994;321197- 1215

Correspondence

CME
Meets CME requirements for:
Browse CME for all U.S. States
Accreditation Information
The American Medical Association is accredited by the Accreditation Council for Continuing Medical Education to provide continuing medical education for physicians. The AMA designates this journal-based CME activity for a maximum of 1 AMA PRA Category 1 CreditTM per course. Physicians should claim only the credit commensurate with the extent of their participation in the activity. Physicians who complete the CME course and score at least 80% correct on the quiz are eligible for AMA PRA Category 1 CreditTM.
Note: You must get at least of the answers correct to pass this quiz.
You have not filled in all the answers to complete this quiz
The following questions were not answered:
Sorry, you have unsuccessfully completed this CME quiz with a score of
The following questions were not answered correctly:
Commitment to Change (optional):
Indicate what change(s) you will implement in your practice, if any, based on this CME course.
Your quiz results:
The filled radio buttons indicate your responses. The preferred responses are highlighted
For CME Course: A Proposed Model for Initial Assessment and Management of Acute Heart Failure Syndromes
Indicate what changes(s) you will implement in your practice, if any, based on this CME course.
NOTE:
Citing articles are presented as examples only. In non-demo SCM6 implementation, integration with CrossRef’s "Cited By" API will populate this tab (http://www.crossref.org/citedby.html).
Submit a Comment

Multimedia

Some tools below are only available to our subscribers or users with an online account.

Web of Science® Times Cited: 67

Related Content

Customize your page view by dragging & repositioning the boxes below.

Articles Related By Topic
Related Topics
PubMed Articles
JAMAevidence.com

Users' Guides to the Medical Literature
Clinical Scenario

Users' Guides to the Medical Literature
Example 1: Diabetes and Target Blood Pressure